MadgeTech, ThermoVault 1-Channel Oven Logger, max. +350°C
The ThermoVault is a thermally insulated, battery-powered, stand-alone, thermocouple-based temperature data logger used for the thermal profiling of
ovens. It is suitable for applications such as powder coating, curing ovens for wet painting, batch or continuous ovens and food processing.
The device records temperature data inside an oven to be later downloaded to the user’s PC.
The ThermoVault will measure and record up to 500,000 measurements. It can withstand an oven temperature of up to +350°C for up to 30 minutes when properly sealed.
Maximum heat exposure time (based on initial temperature of +25°C):
Ambient Temperature: |
Maximum Duration in Minutes: |
+100°C |
136 |
+125°C |
100 |
+150°C |
80 |
+175°C |
66 |
+200°C |
56 |
+225°C |
49 |
+250°C |
44 |
+275°C |
39 |
+300°C |
36 |
+325°C |
32 |
+350°C |
30 |
Its stainless steel thermally insulated enclosure and thermal properties lend itself to be the ideal temperature recording device for applications such as powder coating cure ovens, wet coating cure ovens, batch ovens and conveyor ovens. The supplied external thermocouple probe provides accurate temperature measurements with fast response times.
The device's real time clock ensures that all data is time and date stamped. The storage medium is non-volatile solid state memory, providing maximum data security even if the battery becomes discharged. The device can be started and stopped directly from your computer. The ThermoVault makes data retrieval quick and easy. Plug it in using an optional IFC200 (sold separately) and the user-friendly software does the rest.
Characteristics:
- Temperature data logger with heat protection and type K thermocouple
- Monitoring of extreme temperatures
- Measuring range with supplied sensor: 0 to +482°C
- Oven profiling
- Heat-insulated logger operation temperature range: -20 to max. +350°C (30 minutes)
- Memory for 500,000 readings
- Oven recording interval from 1s
- Battery can be replaced
- Battery service life up to 10 years
MadgeTech 4 Standard, Data Logger Software
This simple, easy-to-use, Windows-based software allows the user to effortlessly collect, display and analyze data. A variety of powerful tools enable checking, exporting and printing professional reports with just one click of the mouse. The latest software can be downloaded free of charge from the MadgeTech website.
- Compatible with the current Microsoft WINDOWS systems
- Record interval from 250ms to 24h
- Immediate start or delayed start up to 6 months
- Date/time stamp
- Display as graph or table
- Data can be exported to MS Excel
- Software languages selectable: English, French, German, Spanish, Russian, ...
Disclaimer and terms of use:
Listed specifications can be used to determine maximum allowable exposure times for the TC101A data logger with ThermoVault heat insulated enclosure. The barrier extends the operating temperature of the logger up to +350°C.
Both the data logger and ThermoVault must be at ambient temperature, approximately +25°C before being placed in the extreme temperature
environment. Immediately following exposure to high temperature, the data logger should be removed from the ThermoVault, using appropriate
precautions, as it could be VERY hot. Failing to remove the data logger may allow heat trapped in the ThermoVault to continue to heat the data logger to potentially unsafe levels.
The ThermoVault may take hours to fully cool down. Even if the exterior of the ThermoVault is cool to the touch, the interior of the barrier and its contents may still be VERY hot. Allow the entire system to cool down to approximately +25°C before successive cycles.
Please note: The TC101A data logger can also be operated without the thermally insulated enclosure. In this case also other thermocouples, thermocouple types with other temperature ranges can be used and the measurements can last extremely longer than with on-site oven measurements.
Technical data:
Sensor technology: |
2 channel temperature
external: Thermocouple type J, K, T, E, R, S, B and N with miniature connector, ≤100 Ohm
internal: semiconductor, ±0.25°C
|
Measuring range: |
from -270°C to +1370°C, depending on the sensor |
Accuracy: |
between ±0.5 °C and ±2.0 °C, depending on the sensor |
Recording interval: |
1s to 24h |
Storage: |
500,000 readings in total
250,000 readings in start/stop mode
|
Interface: |
optional USB interface IFC200 |
Power supply: |
3.6V lithium battery, LTC-7PN,
can be replaced by the user
|
Battery life: |
approx. 10 years @ 15 min interval |
Application area data logger: |
-20ºC to +60 C, 0% RH to 95% RH, non-condensing |
Application Data logger with heat protection: |
-20ºC to +350ºC (30 min), 0% RH to 95% RH, non-condensing |
Dimensions: |
129mm x 113mm x 64mm |
Weight: |
approx. 1590g |
Scope of delivery:
- TC101A, 1-channel thermocouple datalogger
- Thermal insulated enclosure
- Type K thermocouple, 0 to +482°C
- MadgeTech 4 Standard, software via download
- Calibration certificate