Residual Oxygen Measuring System "ResOx 5695-L"
Suitable for measurements in plastic food packaging, with recommended sensor element GOEL 381.
Measuring system with independent battery-operated gas pump for more measuring comfort.
Pressure-compensated, fast measurement, also for rigid packages with low quantities of gas:
- Apply adhesive seal
- Puncture with needle
- Switch on the pump
- After approx. 20s read the minimum value
Application:
Quality control for MAP food packaging and comparable applications.
Technical data:Measuring channels: O
2, T, air pressure
Measuring ranges:
O
2: 0.0 ... 100.0% O
2 or displayed in hPa O
2 / mmHg O
2 Temperature: 0.0 ... 50.0°C
Air pressure: 300 ... 5000hPa (sensor: 500 ... 2000hPa)
Additional functions: Min/Max function - for comfortable measurement of the limit value
Pressure compensation in the gas path - negative pressure in the package/on the sensor is compensated
GOEL 381 sensor: Sensor element: alkaline electrolyte, with special precautions recommended for diving gas and protective gases from 0.0 to 100% O
2, not permanently with increased CO
2 concentration.
Connections on the device::
Sensor: 7-pin bayonet, pressure ports for hoses with inside Ø4mm
Output/ext. power supply: OUT socket: - 38400 baud interface
- Analogue output 0 ... 1V, adjustable
- ext. power supply 5V
Calibration: Quick calibration on air at the push of a button or 2-point/3-point (air +0% and 100%)
GLP: Calibration interval, calibration history
Data logger: Cyclical: 10000, Single: 1000 readings
Single value logger with measuring point entry
Pump: Motorised membrane pump with input/output ports, battery-operated
Max. negative pressure: approx. -360mbar
Delivery rate with GDZ 29 filter: approx. 80ml/min
Connection: Pressure port for hoses with inside Ø4mm
Additional features: Waterproof device and sensor (IP65, IP67), protective armouring, backlighting
Scope of delivery, ready-to-operate system:
Display GMH5695 incl. battery,
Sensor housing with presssure connection incl. sensor GOEL 381,
Gas pump GS150 incl. battery, conection lines, hoses/T-piece,
2 GDZ 29 filters,
2 GOG-N puncture needles Ø0.9mm,
1 GOG-B-B: 45 pcs. adhesive seal,
Carry case GKK 1420